Specifications ZHA-100 DDV
Vessel design life Internal tank s30408, jacket Q245R life 10 years (20000 sterilization cycles)
Design pressure of inner chamber -0.1~0.3MPa
Jacket design pressure 0.3MPa
Built in steam generator (if any) 0.3MPa
Design temperature 150 ̊C
Maximum operating temperature 138 ̊C
Maximum working pressure 0.25MPa
Opening pressure of inner chamber safety valve 0.28MPa
Opening pressure of jacketed safety valve 0.28 MPa
Inner chamber hydrostatic test pressure 0.39MPa
Jacket test pressure 0.52 MPa
Temperature display accuracy 0.1 ̊C
Pressure display accuracy 0.1KPa
Sterilization temperature control accuracy ±02 ̊C
Temperature uniformity ±1 ̊C
Amplitude and times of vacuum pulsation Amplitude -80~0KPa, Frequency 1~99
Number of pulsations Amplitude -80~80KPa, Frequency1
Times of positive 1~3 pressure fluctuation 1~3
Ultimate vacuum -97 KPa
Vacuum leakage rate ≤0.13KPa/min
Surface temperature The external surface temperature does not exceed the ambient temperature by 20 ̊C
Dryness The weight gain of fabrics after sterilization and drying does not exceed 0.6%, and that of utensils, tools and instruments after drying does not exceed 0.1%.
Width × Height 610 × 910
Depth 1815
Volume 1000L
Steam Consumption 55
Length X Width X Height 2120 × 1410 × 2100
Weight 1550kg
Power Supply 220±10% V, 50 Hz