| Specifications |
ZHA-100 DDV |
| Vessel design life |
Internal tank s30408, jacket Q245R life 10 years (20000 sterilization cycles) |
| Design pressure of inner chamber |
-0.1~0.3MPa |
| Jacket design pressure |
0.3MPa |
| Built in steam generator (if any) |
0.3MPa |
| Design temperature |
150 ̊C |
| Maximum operating temperature |
138 ̊C |
| Maximum working pressure |
0.25MPa |
| Opening pressure of inner chamber safety valve |
0.28MPa |
| Opening pressure of jacketed safety valve |
0.28 MPa |
| Inner chamber hydrostatic test pressure |
0.39MPa |
| Jacket test pressure |
0.52 MPa |
| Temperature display accuracy |
0.1 ̊C |
| Pressure display accuracy |
0.1KPa |
| Sterilization temperature control accuracy |
±02 ̊C |
| Temperature uniformity |
±1 ̊C |
| Amplitude and times of vacuum pulsation |
Amplitude -80~0KPa, Frequency 1~99 |
| Number of pulsations |
Amplitude -80~80KPa, Frequency1 |
| Times of positive 1~3 pressure fluctuation |
1~3 |
| Ultimate vacuum |
-97 KPa |
| Vacuum leakage rate |
≤0.13KPa/min |
| Surface temperature |
The external surface temperature does not exceed the ambient temperature by 20 ̊C |
| Dryness |
The weight gain of fabrics after sterilization and drying does not exceed 0.6%, and that of utensils, tools and instruments after drying does not exceed 0.1%. |
| Width × Height |
610 × 910 |
| Depth |
1815 |
| Volume |
1000L |
| Steam Consumption |
55 |
| Length X Width X Height |
2120 × 1410 × 2100 |
| Weight |
1550kg |
| Power Supply |
220±10% V, 50 Hz |