ZPA-250 L

STEAM STERILIZER

Get a Quote
Download Catalog

Description

FEATURES

Specifications ZPA-250 L
Vessel design life Internal tank s30408, jacket Q245R life 10 years (20000 sterilization cycles)
Design pressure of inner chamber -0.1~0.3 MPa
Jacket design pressure 0.3 MPa
Built-in steam generator (if any) 0.3 MPa
Design temperature 150°C
Maximum operating temperature 138°C
Maximum working pressure 0.25 MPa
Opening pressure of inner chamber safety valve 0.28 MPa
Opening pressure of jacketed safety valve 0.28 MPa
Inner chamber hydrostatic test pressure 0.39 MPa
Jacket test pressure 0.52 MPa
Temperature display accuracy 0.1°C
Pressure display accuracy 0.1 KPa
Sterilization temperature control accuracy ±02°C
Temperature uniformity ±1°C
Amplitude and times of vacuum pulsation Amplitude -80~0 KPa, Frequency 1~99
Number of pulsations Amplitude -80~80 KPa, Frequency 1
Times of positive pressure fluctuation 1~3
Ultimate vacuum -97 KPa -97 KPa
Vacuum leakage rate ≤0.13 KPa/min
Surface temperature The external surface temperature shall not exceed the ambient temperature by 20°C
Dryness The weight gain of fabrics after sterilization and drying shall not exceed 0.6%, and that of utensils, tools and instruments after drying shall not exceed 0.1%.
Single-phase voltage supply 220±10% V, 50 Hz
Three-phase voltage supply 380±10% V, 50 Hz
Pure steam 0.25 MPa ~0.35 MPa (Self-produced steam configuration none)
Industrial steam 0.3 MPa ~0.5 MPa (Self-produced steam configuration none)
Water softening 0.15 MPa ~0.3 MPa
Purified water or water for injection 0.15 MPa ~0.3 MPa (Only for self-produced steam configuration)
Compressed air 0.4 MPa ~0.7 MPa